High-throughput multiplexed serology via the mass-spectrometric analysis of isotopically barcoded beads

Main

Serology immunoassays are powerful tools that measure specific antibodies in biological samples. These assays have a wide range of applicability, including monitoring vaccine efficacy1,2,3,4, epidemiological surveillance5, mapping viral transmission dynamics6,7, infectious and autoimmune disease diagnosis8,9, and drug discovery10. In situations with large number of samples requiring assessment across multiple targets11, the scalability of current serology assays is a major bottleneck. There is a pressing need, highlighted by the coronavirus disease 2019 (COVID-19) pandemic, for precise, accurate, highly parallel, low-cost, fast and recurrent serology measurements that use small sample volumes.

Classic immunoassays such as the enzyme-linked immunosorbent assay (ELISA) are widely used because of simplicity, sensitivity and specificity. However, in ELISA, each sample is tested individually for one molecular target at a time, which severely hampers high-throughput applications due to cost and time constraints. In recent years, automation strategies and multiplexed serology techniques have increased throughput12,13,14,15,16,17,18,19,20,21,22,23,24,25. For example, fluorophore barcoded beads used in Luminex assays allow higher throughput by simultaneously detecting multiple analytes in a single sample14,26. However, the spectral overlap of fluorophores constrains the available number of barcodes and, thus, limits the scalability of fluorescent bead-based serology assays. An elegant strategy has recently ameliorated a major issue arising from spectral overlap in beads, nonlinear behaviour of multicolour Förster resonance energy transfer and resulting cascades, but the barcoding space of this method is currently limited to the generation of 580 distinguishable fluorophore combinations27. The maturity and widespread acceptance of bead-based serology assays have allowed these assays to become industry standards. However, scalable paradigms capable of greatly surpassing existing barcoding limitations, to the range of tens of thousands of uniquely barcoded beads, would be a substantial breakthrough for research and clinical applications.

Mass cytometry is a mass-spectrometric, multiparameter and single-cell technology extensively used in the past decade for cell profiling28. Over 50 parameters are routinely analysed on each cell using antibodies conjugated to isotopically enriched rare earth metals28. The instrument provides mass spectra on the fly, one event at a time, at a speed of about 1,000 events per second. In mass cytometry, channel overlap is minimal, and thus, the theoretical barcoding space when 38 stable isotope channels are used is in the range of billions. These unique capabilities of a mass cytometer could be harnessed for high-throughput serology testing by assessing antibody binding to isotopically barcoded beads loaded with specific targets29,30,31. Bead-based assays for cytokine profiling via mass cytometry have been recently reported but are limited to the simultaneous detection of nine analytes32. This mismatch between the theoretical capability of mass cytometry for serology testing and its practical implementation is largely explained by the laborious experimental requirements for the production of isotope-loaded beads compatible with mass cytometry using current techniques32. Therefore, the vast potential of mass cytometry for serology testing has remained largely untapped.

Here, we report a practical, low cost, highly scalable bead-based multiplex serology testing using mass cytometry. We first show that polystyrene beads can be uniformly and robustly loaded postsynthesis with stable isotopes and can be reliably quantified via mass cytometry. We then illustrate the scalability of the method via the generation of 18,480 distinct isotopically barcoded beads. By harnessing this large barcoding space, we report the detection of antibody levels against multiple severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) proteins in multiple clinical serum and plasma samples simultaneously in a one-tube assay. This assay takes one operator approximately 8 h to complete for 924 samples, and no automation equipment is necessary. Our study demonstrates that mass cytometry with barcoded beads facilitates serology testing of multiple antigens in samples from large sample cohorts. This technology has the potential to contribute to management of public health and infectious disease pandemic response.

Results

A postsynthesis strategy for robust incorporation of stable isotopes to polystyrene beads

Two requirements are necessary to enable high-throughput bead-based assays via mass cytometry: (1) uniform incorporation of a high load of isotopes per bead and (2) a scalable strategy generating a large number of isotope-barcoded beads. Dispersion polymerization of isotope-containing polystyrene beads is a standard approach for synthesis of beads containing a high load of isotopes33, but this procedure is not ideal for high-throughput barcode generation because each synthesis is performed individually (Supplementary Fig. 1a). Passive absorption of isotopes postsynthesis of the beads is an appealing strategy for barcoding due to being highly scalable, but in practice, isotope loading is non-uniform across beads, thus impairing data decoding (Supplementary Fig. 1a,b).

We reasoned that binding of isotope-conjugated biotinylated proteins to streptavidin-coated polystyrene beads would provide a high load of isotopes per bead and a uniform labelling, the latter understood as single beads in the population having similar isotope intensities (Fig. 1a and Supplementary Fig. 1a), while being compatible with microfluidic technologies for high-throughput combinatorial dispensing. We conjugated the stable isotope dysprosium 162 (162Dy) to biotinylated bovine serum albumin (BSA) and loaded it to streptavidin-coated beads. Mass cytometry of these beads revealed a high load of isotopes per bead and a uniform labelling (Fig. 1b). Next, we extended the approach to 37 additional stable isotopes, ranging in atomic mass from lanthanum 139 to ytterbium 176 (176Yb) by generating beads with two types of combination of the isotopes. ‘Odd’ beads were loaded with the stable lanthanide isotopes having an odd mass number, and ‘even’ beads were loaded with the stable lanthanide isotopes having an even mass number (Fig. 1c). Using mass cytometry, we observed the expected positive signal for each isotope for both ‘odd’ and ‘even’ beads (Fig. 1d and Supplementary Fig. 1c), and the intensities for each isotope were similar despite the distinct transmission factors for each isotope typically observed in inductively coupled plasma mass spectrometry34 (Fig. 1e and Supplementary Fig. 1d). The analysis of a pool consisting of beads loaded with distinct amounts of biotinylated protein carriers conjugated to 172Yb and 176Yb revealed the expected nine populations of beads (Supplementary Fig. 2), highlighting the tunability, flexibility and robustness of the system. Together, these analyses showed a uniform, tuneable and high load incorporation of 38 stable lanthanide isotopes to polystyrene beads postsynthesis and confirmed that this strategy is compatible with the loading of up to 19 distinct isotopes per bead.

Fig. 1: Postsynthesis strategy for incorporation of stable isotopes on polystyrene beads.
figure 1

a, A schematic representation of streptavidin-coated polystyrene beads loaded with isotope-conjugated biotinylated protein carrier. b, A representative histogram of 162Dy intensity per bead in mono-labelled beads. The 162Dy per bead was measured by mass cytometry. c, A schematic representation of beads loaded with ‘odd’ and ‘even’ isotopes. d, The representative histograms of isotope intensity per bead for ‘odd’ and ‘even’ beads. e, A heat map of isotope intensities for ‘odd’ and ‘even’. The colour shows the median of the normalized isotope intensities for odd and even bead populations.

Full size image

A high-throughput strategy to create thousands of isotope-barcoded polystyrene beads

There are two main components involved in high-throughput serology assay applications: (1) the ability to analyse multiple molecular targets per sample and (2) the ability to analyse large number of samples. To address both, we designed a double-barcoding strategy in which two sets of isotopes are present in each bead: one to identify the molecular target and another for identifying the sample (Extended Data Fig. 1a). Each set of isotopes provides a barcode, and thus, each bead has two barcodes (Extended Data Fig. 1b). Each of the two barcodes consists of k isotopes from a total of n isotopes, where k = n/2 for efficient generation of minimally redundant combinations and robust singlet detection35. The use of two barcodes, instead of a single barcode that encodes both sample and molecular target identification, is also convenient because it enables harnessing postsynthesis bead incorporation of isotopes in sequential loading steps (Extended Data Fig. 1c).

We created 18,480 barcodes using the double-barcode strategy using 18 isotopes, ranging from terbium 159 (159Tb) to 176Yb. Each bead was first loaded with three out of the six isotopes from the first set (159Tb to 164Dy) to generate 20 barcodes, pooled and then loaded with 6 out of the 12 isotopes from the second set (holmium 165 (165Ho) to 176Yb) to generate 924 barcodes (20 × 924 = 18,480 barcodes) (Fig. 2a and Extended Data Fig. 1b). As approximately half of the beads are expected to be positive any given isotope, when analysed, each of the 18 isotopes showed the expected bimodal distribution in the global bead population (Supplementary Fig. 3). We developed an automatic debarcoding pipeline capable of identifying bead events with a correct double-barcode signature (Supplementary Fig. 4). From the analysis of three independent batches of barcoded beads, with a mean of 1.88 ± 0.31 × 106 events per run, we accurately debarcoded 72.9% ± 10.97% of bead events with a correct signature of 6 out of 12 isotopes from the second set, and from those, 85.29% ± 8.91 % of bead events had a correct signature of 3 out of 6 isotopes from the first set (Supplementary Fig. 5). Accurately debarcoded bead events grouped in one of the 20 groups from the first barcode set (Fig. 2b and Supplementary Figs. 6 and 7), and within each group, beads further separated in 924 groups from the second barcode set (Fig. 2c and Supplementary Figs. 8 and 9). Bead counts for each of the 924 barcodes from the second barcode set followed a normal distribution (Supplementary Fig. 10), with 100% recovery of the 18,480 barcodes (Supplementary Fig. 11). Together, these data demonstrated high-throughput generation of thousands of isotope-barcoded beads and their robust identification from a pooled mixture using mass cytometry.

Fig. 2: Double-barcoding strategy enabled generation of 18,480 isotope-barcoded beads.
figure 2

a, A schematic representation of the workflow for the generation of isotope-barcoded beads. For target ID barcoding, the beads are labelled with 3 of the 6 target ID isotopes (159Tb to 164Dy) to generate 20 barcodes (BCs). For sample ID barcoding, the beads are labelled with 6 of the 12 sample ID isotopes (165Ho to 176Yb) to generate 924 BCs. b, Left: a uniform manifold approximation and projection (UMAP) of beads grouped by the intensity of the target ID isotopes and coloured by target ID BC. Right: representative groups coloured by the intensity of the target ID isotopes. Each dot represents a bead. The data are from three independent experiments. c, Left: a UMAP of beads from BC 20 label in b grouped by the intensity of the sample ID isotopes and coloured by sample ID BC. Right: representative BCs 549 and 619 coloured by the intensity of the sample ID isotopes. Each dot represents a bead. The data are from three independent experiments.

Full size image

Identification of thousands of barcoded beads from a mixture offers a versatile system for the design of different types of multiplexed serology assay with unique capabilities. As a demonstration of these capabilities, we designed two systems: the first for detection of serum antibodies using a very low sample volumes (Fig. 3) and the second for rapid, wash-free, high-throughput detection of serum antibodies (Fig. 4).

Fig. 3: Two-bead strategy to detect specific antibodies in low volumes of plasma.
figure 3

a, A schematic representation of the two-bead strategy for detection of antibodies in low volumes of plasma. b, A schematic representation of the baseline and flowthrough. The beads bound by anti-IgG magnetic beads are retained in the column, and the unbound beads are collected in the flow through. c, A dot plot of the number of barcoded beads in the baseline versus number in the flowthrough for 20 serum samples collected before the COVID-19 pandemic (green) and 20 samples from patients who had tested positive for COVID-19 (red). d, A comparison of ELISA (y axis) versus the two-bead assay (x axis) in the 20 COVID-19-positive samples shown in c. The two-bead assay result is shown as the ratio of the number of beads in the flowthrough to the number of beads in the baseline. Note that the two samples in c that behave as prepandemic samples also show low ELISA values. e, A two-bead assay on a set of 39 COVID-19-positive samples and 55 prepandemic samples. The dot colour indicates the sample volume: 100 nl is red, 10 nl is green and 1 nl is blue. Each dot is the mean of 12 replicates (3 sample volume replicates and 4 target replicates per sample volume). The targets were: COVID-19 spike S1, spike S1 RBD, spike S2, nucleocapsid and negative (no target). f, A heat map of the values of the 12 replicates for each red dot shown in e.

Full size image
Fig. 4: A high-throughput and wash-free strategy for multiplexed detection of antibodies.
figure 4

a, A schematic representation of the wash-free strategy to detect antibodies against 20 targets in 924 samples in a single mass cytometry analysis. b,c, A heat map of IgG (b) and IgM (c) levels against 19 targeted proteins (rows) in 542 samples (columns) from the FIND study, which collected samples from individuals with confirmed SARS-CoV-2-positive tests along with additional household members. The samples are labelled by day of sample collection after positive symptoms for the household’s index case (timepoint), PCR test and vaccination status. IgG and IgM levels were normalized to values of beads not loaded with a target antigen (negative). The samples and targets are hierarchically clustered by IgG or IgM levels. d, A boxplot of the IgG levels against the SARS-CoV-2 spike trimer per sample by timepoint. Each dot represents a sample. e, A dot plot of the levels of IgG against the SARS-CoV-2 spike S1 trimer obtained by mass cytometry versus those obtained using the Abbott AdviseDx SARS-CoV-2 IgG II assay. Each dot represents a sample. The orange line is a polynomial regression line of degree 2. R2 = 0.58. f, A dot plot of the levels of IgG against the beta SARS-CoV-2 spike trimer versus the wild-type trimer quantified using mass cytometry. Each dot represents a sample. NTD, N-terminal domain.

Full size image

An assay for detection of plasma antibodies using low sample volume

We designed a two-bead selection system to detect the presence of specific antibodies in low volumes of human plasma (Fig. 3a). In this system, one bead is barcoded with a predefined combination of stable isotopes and loaded with a biotinylated antigen of interest (for example, a purified protein or a protein subunit). The other bead is magnetic and is conjugated to antihuman immunoglobulins. Cross-linking of barcoded beads to magnetic beads occurs only in the presence of antibodies recognizing target antigens (Extended Data Fig. 2). For example, in an assay leveraging the two-bead selection system, each clinical plasma sample is incubated in individual wells with barcoded beads loaded with distinct antigens to allow host antibodies to bind to antigens on beads. The barcoded beads are then washed to remove any unbound antibody and pooled across all samples to create a baseline bead mixture. The magnetic beads are then added providing an opportunity for cross-linking of magnetic beads to barcoded beads via antihuman immunoglobulins when host antibodies are present. Washes are performed using a magnet to retain only those barcoded beads from host antibody positive samples, and the flowthrough is then analysed by mass cytometry. For each barcode, the ratio of the bead count in the flowthrough and the baseline bead mixture provides an estimate of host antibody levels, with low ratio values implying the presence of host antibodies specific to the antigen in a given sample.

As a demonstration of this method, we first incubated barcoded beads loaded with the recombinant SARS-CoV-2 S1 subunit of the spike protein (spike S1) with 20 plasma samples from COVID-19 convalescent donors and 20 negative control plasma samples collected before the emergence of COVID-19, all diluted 300-fold. The 40 distinct types of barcoded bead were similarly represented in the pooled baseline mixture, which is collected before magnetic bead selection (Fig. 3b,c). Consistent with expected patterns of serum immunity, after selection, all barcoded beads incubated with pre-COVID-19 pandemic samples were enriched (Fig. 3b,c), and most barcoded beads incubated with patient samples positive for SARS-CoV-2 were depleted (Fig. 3c). Two of the SARS-CoV-2-positive samples resembled pre-COVID-19 samples (Fig. 3c). Consistently, independent ELISA analysis of the SARS-CoV-2-positive samples showed high anti-spike S1 levels for all samples except the two behaving similarly to pre-COVID-19 samples (Fig. 3d), suggesting these two patient samples may have low levels of SARS-CoV-2 antibodies despite being from individuals with prior SARS-CoV-2-positive polymerase chain reaction (PCR) tests. We observed similar results in three independent experiments (Supplementary Fig. 12). Furthermore, we observed similar results when samples were analysed individually in an independent flow cytometry assay rather than pooled together in the mass cytometry assay (Supplementary Fig. 13a). We additionally observed proper alignment of a multiplexed assessment of the levels of antibodies against SARS-CoV-2 spike S1, the receptor binding domain of spike S1 (spike S1 RBD), and the nucleocapsid protein to the results of an ELISA assay (Supplementary Fig. 13b). To gauge sensitivity of the assay, we incubated beads loaded with spike S1 with increasing concentrations of a monoclonal antibody against spike S1 and observed detection down to 100 pg ml−1 for this particular antibody-antigen interaction (Supplementary Fig. 14). These results show that the mass cytometry results align with the results obtained with ELISA and flow cytometry, validating the two-bead system for antibody detection in patient plasma samples.

We next leveraged the extensive barcoding space of the isotope-tagged beads to analyse an additional set of 39 plasma samples from individuals who had tested positive for SARS-CoV-2 by PCR and 55 plasma samples collected before the COVID-19 pandemic. Each sample was diluted 1:300, 1:3,000 and 1:30,000, which is equivalent to using 100, 10 and 1 nl of sample, respectively, in a 30 μl reaction volume. We incubated the samples with barcoded beads loaded with recombinant SARS-CoV-2 spike S1, spike S1 RBD, the S2 subunit of the spike protein (spike S2) and the nucleocapsid protein. Each sample and dilution were quantified 12 times using three sample replicates and four target replicates by calculating the ratio of the bead count in the flowthrough and the baseline bead mixture in each replicate (Extended Data Fig. 3). The two-bead selection system was able to distinguish all 39 patients with COVID-19 samples from negative controls with 100% success rate using 100 nl of sample volume (Supplementary Fig. 15). In some samples, antibodies against spike S1 were detected using only 1 nl of sample (Fig. 3e; samples 21 and 36, spike S1, blue dots). The assay also revealed patient-specific variability in SARS-CoV-2 antibody profiles. For instance, some samples had high levels of antibodies against spike S1 but not against spike S1 RBD, spike S2 or the nucleocapsid (Fig. 3e; samples 14 and 23, red dots). Others had high levels of antibodies against spike S1, spike S1 RBD, spike S2 and nucleocapsid (Fig. 3e; samples 16 and 28, red dots). Lastly some had high levels of antibodies against spike S1 and nucleocapsid but not spike S1 RBD and spike S2 (Fig. 3e; samples 13 and 15, red dots). For each sample, each of the 12 replicates had similar values (Fig. 3f and Supplementary Fig. 16), and the results correlated well in two independently performed assays (Supplementary Fig. 17a). In addition, the levels of antibodies against SARS-CoV-2 spike S1 and nucleocapsid had a strong correlation with independent ELISA assays conducted on the same set of samples (Supplementary Fig. 17b). These data show that a two-bead strategy can detect host antibodies in low volume samples, down to 1 nl, and can provide unique profiles of the patient’s antibody responses against multiple antigens.

A high-throughput, wash-free assay for multiplexed detection of antibodies

Secondary antibodies are advantageous for serology assays due to ease of use and have been previously used in low-throughput bead-based mass cytometry assays32. We validated the compatibility of secondary antibodies conjugated to isotopes for the quantification of host immunoglobulins bound to the isotope-barcoded beads using positive and negative controls. Spike S1-loaded, isotope-barcoded beads were incubated with dilutions of a plasma sample from an individual positive for SARS-CoV-2 antibodies and with plasma samples from eight individuals collected before the COVID-19 pandemic as negative controls. The beads were washed, pooled and incubated with antihuman IgG conjugated to gold nanoparticles. We observed a positive control signal well above the background (Supplementary Fig. 18, sample 16 versus samples 1–12) that spanned linearly from 8 µl to 62.5 nl in a 1:2 serial dilution series (Supplementary Fig. 18, samples 13–20). We observed minimal non-specific binding of the secondary anti-IgG antibody to beads (Supplementary Fig. 18, samples 1–4) and low non-specific binding of prepandemic samples (Supplementary Fig. 18, samples 5–12). These results validated that antibodies bound to antigen-loaded, isotope-barcoded beads can be readily detected via secondary antibodies.

A major factor limiting the scalability of serology assays such as ELISA and fluorescently barcoded bead-based assays is the need to perform multiple washes, one plate at a time, to remove unbound antibodies. Automation strategies can ameliorate the issue36 but are not straightforward to implement. We designed a system consisting of consecutive fixation and denaturation steps to eliminate washing constraints for isotope-barcoded beads serology assays (Fig. 4a). In this wash-free system, antigen-loaded isotope-barcoded beads are first incubated with samples of interest to allow host antibodies to bind to bead loaded with antigens. Each well is then treated with paraformaldehyde (PFA) to fix host antibodies onto the beads. A subsequent sodium dodecyl sulphate (SDS) treatment denatures the bound and unbound immunoglobulins (Supplementary Fig. 19) without removing isotopes from the beads (Supplementary Fig. 20). The samples are all then pooled and washed. Pooled samples are then incubated with isotopically labelled secondary antibodies, and the analysis is completed by mass cytometry to reveal the level of bound host antibodies on each barcoded bead.

To demonstrate bead-based multiplex serology testing by mass cytometry at scale in a wash-free system, we simultaneously analysed the levels of IgG (Fig. 4b) and IgM (Fig. 4c) against 19 distinct SARS-CoV-2 protein variants (Supplementary Table 1) in 542 longitudinal serum samples collected from a cohort of SARS-CoV-2 infected households in San Francisco (Supplementary Table 2). As part of this study blood samples were collected longitudinally from index cases with positive SARS-CoV-2 tests and additional household members. Extensive epidemiological data were also captured that included prior SARS-CoV-2 vaccination status. Expectedly, anti-wild-type SARS-CoV-2 Trimer IgG levels increased over time across the cohort as a whole (Fig. 4d), and these increased levels were strongly related to both vaccination status and presence of a positive SARS-CoV-2 PCR test (Supplementary Fig. 21). In the same assay, we included triplicates or quadruplicates of 100 negative samples collected before the COVID-19 pandemic. The immunoglobulin levels against wild-type SARS-CoV-2 trimer were below a given threshold in 97.5% of the tests for IgG and in 100% of the tests for IgM (Supplementary Fig. 22). In the samples collected at day 28 from infected and vaccinated individuals of the San Francisco study, we observed that all 23 samples were above the threshold for IgG (Supplementary Fig. 23), and 17 of the 23 samples were above the threshold for IgM (Supplementary Fig. 24). The levels of IgG against the wild-type SARS-CoV-2 trimer obtained via mass cytometry correlated well with those obtained using Abbott AdviseDx SARS-CoV-2 IgG II (Architect) (Fig. 4e), an assay that received emergency use authorization by the Food and Drug Administration.

Because this assay provides internally controlled data on multiple targets for the same sample, it is feasible to make robust and direct comparisons within individuals across SARS-CoV-2 variants. We observed a high correlation on the antibody response against SARS-CoV-2 variants for most samples (Fig. 4f) but, interestingly, identified a subset of samples with unique responses to certain variants (Fig. 4f) that could reflect variation in antibody repertoire driven by patient or virus-related factors. For example, sample numbers 4,567, 4,582 and 4,599 showed lower IgG levels against the beta, gamma, delta and omicron variants compared with the wild-type and alpha variant, and sample number 4,642 showed lower IgG levels against the beta, gamma and omicron variants compared with the wild-type, alpha and delta variants (Fig. 4f and Supplementary Fig. 25).

These data show that bead-based multiplex serology testing by mass cytometry is quantitative, sensitive and specific. This assay is capable of generating data comparable to existing authorized tests but with massive improvements in scalability, performing 36,960 tests in one tube using 400 nl of sample volume in a 30 μl reaction volume.

Discussion

We report a highly scalable and multiplexed serology technology with readout via mass cytometry. We share the generation of 18,480 unique barcodes by incorporating distinct combinations of stable isotopes on polystyrene beads after synthesis. We also provide an automatic pipeline to debarcode the raw data without the need to input user-defined parameters. Using this technology and a bead-selection approach, we show the detection of IgGs against SARS-CoV-2 spike S1 using the equivalent of 100 nl of sample in a cohort of 55 prepandemic and 39 COVID-19 samples. In a subset of those samples, we show detection of IgGs using the equivalent of 10 and 1 nl of sample. Using secondary detection antibodies, we demonstrate the simultaneously analysis of IgG and IgM levels against 19 proteins in 924 samples in an assay that took one operator approximately 8 h to complete without the need of automation equipment. This is the equivalent of performing 36,960 ELISA-style serology assays in a single run. The ability to run tens of thousands of assays in parallel makes it a cost-effective technology (Supplementary Note 1). Moreover, compared with other strategies used in diagnostic labs such as ELISA and Luminex assays, the hands-on time of operators running the same number of assays is substantially reduced.

The assessment of humoral immunity in the FIND cohort allowed benchmarking of the serology assay via mass cytometry versus an immunoassay approved for clinical use. The data indicate that the mass cytometry assay can detect unique SARS-CoV-2 variant specific signatures in a robust fashion. This detection is made possible due to internal control resulting from pooling and simultaneous processing of all samples. This serology technology is immediately extensible into additional antigen variants from SARS-CoV-2 (ref. 37) and other targets by producing the required recombinant proteins. Comprehensive characterization of immunoglobulin subclasses coupled to longitudinal testing campaigns can reveal immune mechanisms that might inform vaccination campaigns, as illustrated by a preferential class switch to IgG4 after booster shots of an mRNA vaccine for SARS-CoV-2 (ref. 38). A future development could harness part of the remaining isotope space to include simultaneous assessment of IgG1, IgG2, IgG3, IgG4, IgM, IgA1 and IgA2 on each bead for deeper sample profiling. Other antibody locations with key biological relevance, such as the nasal mucosa39 or gingival crevice40, could be also interrogated by expanding the assay to additional sample sources.

From a technical standpoint, the current iteration of the isotope-based serology assay is mainly constrained by the number of available barcodes and acquisition speed. Over 50 isotopes are available off-the-shelf for mass cytometry28, and we report the simultaneous use of 20 of them in the serology assay. The remaining isotopes could be harnessed to increase the number of barcodes, but careful validation should be performed given the distinct transmission factors for each isotope in inductively coupled plasma mass spectrometry. For instance, using 16 instead of 12 isotopes for sample ID would generate 12,870 combinations, and 8 instead of 6 isotopes for target ID would generate 70 combinations, for an aggregate of 900,900 barcodes. An increase in the number of barcodes should also be accompanied by further assay miniaturization and automation. The inductively coupled plasma time-of-flight mass spectrometry instrument used in this report has a relatively slow sample acquisition speed of 1,000 events per second, which limits the number of samples that can be analysed, but a higher speed could be achieved using alternative instruments41, smaller particles and multiplex signal amplification systems42,43. The beads used in this study have a diameter variation of 3–3.4 µm, with larger beads having a higher number of streptavidin molecules on their surface. The use of beads with a narrower range of diameter variation will also improve the throughput of the assay by reducing the number of required beads for analysis. Moreover, the assembling of antigen panels is constrained by the scalability of protein purification. Our technology scales faster than the process of integrating new purified proteins into the panel, thus posing a challenge in fully leveraging the multiplexing potential of the technology. The commercial availability of antigens ameliorates the issue, although extensive validation will be required when assembling new antigen panels.

In summary, we have designed and implemented a fast, high-throughput and cost-effective serology technology that opens new opportunities for profiling humoral immune responses at population scales. We envision that this technology will have a broad range of eventual applications, for instance in government-sponsored serology surveillance, drug-discovery efforts, monitoring of vaccination campaigns, diagnosis of autoimmune diseases and cancer and monitoring of immune correlates of protection during infectious disease outbreaks. These applications of mass cytometry-based serology will help reveal humoral immune mechanisms at a systems level, which may guide development of therapeutic strategies and enable prediction of clinical outcomes.

Methods

Clinical samples

Deidentified samples from the study44 were used for Fig. 3c,d, comprising individuals with confirmed SARS-CoV-2 infection through PCR with reverse transcription (RT–PCR) testing. The samples used here were collected from COVID-19 convalescent plasma donors who donated plasma at the Stanford Blood Center between April 2020 and May 202044. This study was approved by the Stanford University Institutional Review Board (protocols IRB-13952, IRB-48973 and IRB-55689).

Deidentified COVID-19-positive samples used for Fig. 3e,f were purchased from Discovery Life Sciences, comprising individuals with confirmed SARS-CoV-2 infection through RT–PCR testing.

Deidentified samples from FIND COVID were used for Fig. 4. FIND COVID is a Centers for Disease Control and Prevention (CDC)-funded longitudinal cohort of individuals recently diagnosed with SARS-CoV-2 infections in the San Francisco Bay Area, initiated in August 2020. The index cases were identified from individuals with a positive health provider-ordered SARS-CoV-2 nucleic acid amplification test at a UCSF-affiliated health facility. Index cases were enroled with uninfected household contacts and followed at five field visits over the first 4 weeks of illness. Study visits occurred on the day of enrolment and days 9, 14, 21 and 28 postindex symptom onsets. The blood samples were collected form index cases and household contacts at each study visit. Moreover, the participants self-collected anterior nasal swabs daily for the first 2 weeks following enrolment, then on days 17, 19, 21 and 28. All nasal specimens were used to quantify SARS-CoV-2 RNA through RT–PCR targeting nucleocapsid (N) and envelope (E) genes to determine SARS-CoV-2 infection status of household participants. The study was approved by the UCSF Institutional Review Board (protocol IRB number 20-30388) and given a designation of public health surveillance according to federal regulations as summarized in 45 CFR 46.102(d)(1)(2).

Deidentified prepandemic samples used for Figs. 3 and 4 were purchased from Solomon Park, comprising samples collected before November 2019 (the official start of the COVID-19 pandemic).

Conjugation of BSA and antibodies to isotopes

Biotinylated BSA (Sigma, number A8549) was used as carrier protein and it was conjugated to isotope-chelated Maxpar X8 polymers (Fluidigm, number 201300). Briefly, 200 μg of biotinylated carrier proteins were reduced by tris(2-carboxyethyl)phosphine (TCEP) treatment (Thermo Fisher Scientific, number 77720) for 30 min at 37 °C. The reduced carrier proteins were reacted with isotope-chelated maleimide-containing polymers for 1.5 h at 37 °C and washed five times with 1× PBS. The protein concentration was quantified using a Nanodrop 2000 (Thermo Fisher Scientific, number ND-2000) and all conjugates were diluted to 0.5 g l−1 in 1× PBS. Isotope conjugation was confirmed after every conjugation experiment by loading isotope-conjugated BSA into beads and analysing them by mass cytometry. Antihuman IgM was conjugated to 115In chelated Maxpar X8 polymers following the same procedure.

Bead loading with isotope-conjugated carrier proteins

Streptavidin-coated polystyrene beads with a diameter of 3–3.4 μm (Spherotech, number SVP-30-5) were washed thrice in 1× PBS with 0.5% BSA and 0.001% Tween-20 (CSM-T). The beads where then incubated with 2.5 μg ml−1 isotope-conjugated biotinylated carrier proteins for 30 min at room temperature, washed thrice in CSM-T and stored at 4 °C.

Bead loading with isotopes by passive absorption

Streptavidin-coated polystyrene beads with a diameter of 3–3.4 μm (Spherotech, number SVP-30-5) were washed thrice in CSM-T. The beads where then incubated with 50 μM lanthanide chloride in 1× PBS for 1 h at room temperature, washed thrice in CSM-T and stored at 4 °C until analysis.

Load proteins on beads for 924 assays

For 1 ml of streptavidin-coated polystyrene beads with a diameter of 3–3.4 μm (number SVP-30-5, Spherotech), 10 μl CSM with 0.1% Tween-20 (Sigma, number P1379) were added. The beads were washed thrice with 500 μl CSM-T by centrifuging at 2,800 relative centrifugal force (RCF) at room temperature. The supernatant was discarded. For each avi-tagged protein a 1:10 dilution was prepared (stock at 0.25 g l−1) by adding water. The stock was quickly stored at −20 °C after usage. The beads were resuspended in 165 μl CSM-T and then split in 20 tubes (7.5 μl each) labelled from 1 to 20. To each tube, 7.5 μl of an avi-tagged protein dilution was added to the beads. The beads with the avi-tagged protein were incubated for 30 min at room temperature. The beads were washed thrice with 100 μl CSM-T by centrifuging at 2,800 RCF at room temperature. To each of the tubes 165 μl of CSM-T was added. The peads loaded with avi-tagged proteins were store at 4 °C until use.

Generation of 20 barcodes for target ID

To create 20 barcodes, three isotope-conjugated biotinylated carrier proteins from a collection of six were distributed in wells of ten V-bottom 96-well plates (Costar, number 3363) containing 94 μl of CSM-T. The concentration of each individual three isotope-conjugated biotinylated carrier proteins was 1 μg ml−1. After dispensing, 50 μl of the beads loaded with avi-tagged proteins were placed on each well. The plates were incubated for 30 min at room temperature. The beads were then washed thrice in CSM-T by centrifugation at 2,000 RCF at room temperature and stored at 4 °C until use.

Generation of 924 barcodes for sample ID

High-throughput generation of barcodes was performed on a Tempest (Formulatrix). To create 924 barcodes, 6 isotope-conjugated biotinylated carrier proteins from a collection of 12 were distributed in wells of ten V-bottom 96-well plates (Costar, number 3363) containing 44 μl of CSM-T. After dispensing, the concentration of each individual six isotope-conjugated biotinylated carrier proteins was 1 μg ml−1. The plates were manually transferred to a Bravo Automated liquid Handler Platform (Agilent) and 50 μl of the beads loaded with avi-tagged proteins were placed on each well. The plates were incubated for 30 min at room temperature. The beads were then washed thrice in CSM-T by centrifugation at 2,000 RCF at room temperature and stored at 4 °C until use.

Two-bead assay by flow cytometry

For each sample, 5 μl of fluorescently labelled streptavidin polystyrene beads (Spherotech, numbers SVFP-0556-5 and SVFP-1068-5) were washed twice with CSM-T at a 1:1 ratio and pelleted by centrifuging at 21,000 RCF for 2 min after each wash and discarding the supernatant. Next, the beads were resuspended in 5 μl CSM-T per sample. Then, per sample, 1 μl of biotinylated spike S1 at 0.25 μg μl−1 (Sino Biological, number 40591-V27H-B) was added to 5 μl of beads and was incubated for 30 min at room temperature. The beads were washed three times by adding 100 μl of CSM-T, centrifuging at 21,000 RCF for 3 min and discarding the supernatant. Then the pellet was resuspended in 30 μl of CSM-T and transferred to a 96-well V-bottom plate (Costar, number 3363). Next, 10 μl of 1:10 diluted plasma in CSM-T (or plasma titrated from 1, to 1:1,000 μl) or an antibody titration was added to each condition. The beads were incubated for 30 min at room temperature and washed twice by adding 50 μl of CSM-T, centrifuging at 2,800 RCF for 3 min and discarding the supernatant. The beads were then resuspended in 100 μl of CSM-T. In each well of the plate, 2 μl of Magnetic-IgG beads (RayBiotech, number 801-101-1) were added. The beads were incubated for 30 min at room temperature with mixing by pipetting every 10 min. Subsequently, another 100 μl of CSM-T was added to each well and the plate was placed on a 96-well magnetic separator plate for 2 min to attach the magnetic beads. The unbound beads in solution were collected and added to another well. The collected beads were washed three times by adding 100 μl of CSM-T, centrifuging at 2,800 RCF for 3 min and discarding the supernatant. Finally, the beads were resuspended in 100 μl of CSM-T and analysed using a CytoFlex flow cytometer (Beckman Coulter). For each sample, acquisition was based on a specific volume, either 20 or 50 µl. Fluorescent Blue and Nile Red beads were initially identified and gated using forward and side scatter channels and subsequently gated on the basis of their fluorescence signal. This volume-based acquisition approach enabled the counting of the total number of beads in each sample volume. The counts obtained were then analysed and plotted using R.

Two-bead assay by mass cytometry

The beads from the barcoded 96-well V-bottom plates (Costar, number 3363) loaded with biotinylated spike S1 (Sino Biological, number 40591-V27H-B) as described above were pelleted by centrifuging at 2,800 RCF, and the supernatant was discarded and then resuspended in 30 μl of CSM-T. Patient plasma was diluted in CSM-T 1:10, and 10 μl of each sample was mixed with the barcoded beads. The plate was incubated for 30 min at room temperature. The beads were washed twice by adding 100 μl of CSM-T, centrifuging at 2,800 RCF for 3 min and discarding the supernatant. Then, 100 μl of diluted spike S1 protein (2 ng μl−1) in CSM-T (Sino Biological, number 40591-V02H) was added to each well and incubated for 10 min at room temperature. The plate was centrifuged at 2,800 RCF, the supernatant was discarded and another 50 μl of spike protein (0.2 ng μl−1) in CSM-T was added. All samples were combined in a reservoir and added to a 50 ml tube and spun down at 2,800 RCF. Then, 10% of the beads were collected and labelled as ‘baseline’. The samples were then diluted in 2.5× the number of samples in CSM-T (for 40 samples, we add 100 μl of CSM-T). We then added magnetic-activated cell sorting anti-IgG microbeads (Miltenyi Biotec, number 130-047-501) to a 1:2 ratio and incubated at room temperature for 30 min. The magnetic-activated cell sorting LS columns (Miltenyi Biotec, number 130-042-401) were prewashed with 1 ml CSM-T. The samples were diluted ten times and added to a LS column—1 ml of diluted sample was added to one LS column—and flowthrough was collected in a 5 ml protein-low bind tube (Fisher Scientific, number 0030122356). Another 4 ml was added and the flowthrough was collected. The beads from the flowthrough were pelleted at 2,800 RCF, and the supernatant was discarded. The beads were resuspended in 1 ml CSM-T and then then moved to a 1.5 ml protein-low bind tube (Fisher Scientific, number 022431081) and pelleted at 21,000 RCF. The beads were washed by adding 1 ml of CSM-T, centrifuging at 21,000 RCF for 3 min and discarding the supernatant. The beads were then washed three times with 0.5 ml H2O before loading into CyTOF. The beads were analysed using a CyTOF 2 mass cytometer (Fluidigm) at a rate of 500 events per second. The focus of this analysis was on isotopes within the mass range of 159–176 atomic mass units.

Wash-free assay by flow cytometry

The beads in 96-well V-bottom plates (Costar, number 3363) loaded with biotinylated spike S1 (Sino Biological, number 40591-V27H-B) were pelleted by centrifuging at 2,800 RCF, and the supernatant was discarded and the then resuspended in 30 μl of CSM-T. Spike S1 was detected by a human anti-spike S1 (Invivogen, number srbd-mab1) or patient plasma diluted 1:10 and 10 μl of each sample was mixed with the loaded beads. The plate was incubated for 30 min at room temperature. Next, 40 μl of PFA (3.2%) was added to each well of the respective samples and incubated for 15 min. Then, 80 μl of 4% SDS was added and incubated for 15 min. The plate was centrifuged at 2,800 RCF, and the supernatant from the beads was discarded. The wells were washed three times by adding 150 μl of CSM-T, centrifuging at 2,800 RCF for 3 min and discarding the supernatant. Next 5 μl of (1:200 dilution) of anti-IgG-647 (Thermo Fischer Scientific, number A21445) was added to each well. The beads were incubated at room temperature for 15 min. The wells were then washed three times by adding 150 μl of CSM-T, centrifuging at 2,800 RCF for 3 min and discarding the supernatant. Each well was resuspended in 150 μl of CSM-T and analysed by a CytoFlex flow cytometer (Beckman Coulter). For each sample, a volume of 50 µl was analysed. The beads were first distinguished and gated using forward and side scatter channels, followed by analysis based on the fluorescence intensity of anti-IgG conjugated with Alexa Fluor 647. Subsequent data analysis was performed utilizing R software.

Wash-free assay by mass cytometry

The beads from the barcoded 96-well V-bottom plates (Costar, number 3363) loaded with biotinylated spike S1 (Sino Biological, number 40591-V27H-B) or 19 distinct SARS-CoV-2 protein variants (Supplementary Table 1) were pelleted by centrifuging at 2,800 RCF, and the supernatant was discarded and the then resuspended in 30 μl of CSM-T. Patient plasma were diluted 1:10, and 10 μl of each sample was mixed with the loaded beads. The plate was incubated for 30 min at room temperature. Next, 40 μl PFA (3.2%) was added to each well of the respective samples and incubated for 15 min. Then, 80 μl of 4% SDS was added and incubated for 15 min. Then, all the samples were combined in a reservoir by pipetting and then added to a 50 ml tube. The beads were washed three times with by adding 25 ml of CSM-T, centrifuging at 2,800 RCF for 3 min and discarding the supernatant. Next, the beads were stained with 25 ml CSM-T containing 256 ng l−1 gold anti-IgG antibody (Nanoprobes, number 2053) and 256 ng l−1 115In-conjugated anti-IgM antibody (BioLegend, number 314502) for 30 min. The beads were pelleted at 2,800 RCF, resuspended in 200 μl CSM-T and transferred to a protein-low bind Eppendorf tube (Fisher Scientific, number 022431081). The samples were washed thrice with 200 μl of CSM-T and pelleted by centrifuging at 21,000 RCF after each wash and discarding the supernatant. The beads were then washed three times with 0.5 ml H2O before loading into CyTOF. The beads were analysed using a CyTOF 2 mass cytometer (Fluidigm), at a rate of 500 events per second. The focus of this analysis was on isotopes within the mass range of 159–176, 115 and 197 atomic mass units. A step-by-step illustration of the assay and a document with the numbers to control and optimize the stoichiometry of biotin to streptavidin at the population level are described in Supplementary Notes 2 and 3.

ELISA to detect anti-SARS-CoV-2 antibodies in plasma samples

The ELISA protocol performed for Fig. 3d and Supplementary Fig. 13b in this study was previously described44,45. In short, 96-well Corning Costar high binding plates were coated with SARS-CoV-2 spike S1, RBD or N protein in PBS at a concentration of 0.1 μg per well overnight and at 4 °C. The plates were then blocked using PBS–Tween-20 containing 3% non-fat milk powder for 1 h. The plasma samples from COVID-19 convalescent plasma donors were incubated at a dilution of 1:100 in dilution buffer (PBS-T containing 1% non-fat milk powder) for 1 h. Anti-SARS-CoV-2 IgG antibodies were detected using horseradish peroxidase-conjugated goat antihuman IgG (Thermo Fisher, number 62-8420; 1:5,000 dilution). The plates were developed using 3,3′,5,5′-tetramethylbenzidine. The reaction was stopped by adding sulfuric acid after 12 min. The optical density was measured at 450 nm, and blank control values were subtracted. All the samples were tested twice in independent experiments. For Supplementary Fig. 17b, we used the ELISA kits: RAPID MAX SARS-CoV-2 Spike S1 Human IgG ELISA Kit (BioLegend, catalogue number 447809) and LEGEND MAX SARS-CoV-2 Nucleocapsid Human IgG ELISA Kit (BioLegend, catalogue number 448107). The protocol was implemented in strict accordance with the manufacturer’s instructions with 500 nl of plasma sample.

Abbott AdviseDx SARS-CoV-2 IgG II assay

SARS-CoV-2-specific antibodies were determined using the immunoassay AdviseDx SARS-CoV-2 IgG II for qualitative and semiquantitative detection of IgG antibodies to the Spike S1 RBD protein according to the manufacturer’s specifications. Serologic analyses for this study using the Abbott AdviseDx SARS-CoV-2 IgG II (Architect) were performed under protocols approved by the UCSF Institutional Review Board (IRB number 11-05519).

Automatic bead debarcoding

The beads were debarcoded using a custom R script. Briefly, all flow cytometry standard data files of a CyTOF run were concatenated. The events were then arcsine transformed with a cofactor of 5 and scaled using min-max normalization. The data for each isotope were plotted in a histogram to confirm a bimodal distribution. The signal intensities for each isotope were automatically thresholded using auto_thresh of the autothresholdr R package (version 1.3.2) with IJDefault method. A vector of thresholded intensities was used to transform the detected events to a binary matrix, in which ‘0’ means signal lower than the threshold and ‘1’ means signal higher than the threshold. Target and sample IDs were assigned following the target ID key (Supplementary Table 3) and sample ID key (Supplementary Table 4). The events with target and sample ID were selected for downstream analysis.

Data visualization

The plots were created using the ggplot2 R package46. For the boxplots: the centre is the median, the minima and maxima bound of box are the 25th and 75th percentiles, respectively, the whiskers extend from the minima and maxima bounds of box to the largest value no further than 1.5 times the interquartile range, and the outliers are shown as dots. Figures 1a, 3a and 4a, Extended Data Figs. 1–3, Supplementary Figs. 1a, 2a, 3, 4a and 18a, and Supplementary Note 2 were created in part using BioRender.com. All figures were prepared using Illustrator (Adobe).

Reporting summary

Further information on research design is available in the Nature Portfolio Reporting Summary linked to this article.

Related Articles

Type 2 immunity in allergic diseases

Significant advancements have been made in understanding the cellular and molecular mechanisms of type 2 immunity in allergic diseases such as asthma, allergic rhinitis, chronic rhinosinusitis, eosinophilic esophagitis (EoE), food and drug allergies, and atopic dermatitis (AD). Type 2 immunity has evolved to protect against parasitic diseases and toxins, plays a role in the expulsion of parasites and larvae from inner tissues to the lumen and outside the body, maintains microbe-rich skin and mucosal epithelial barriers and counterbalances the type 1 immune response and its destructive effects. During the development of a type 2 immune response, an innate immune response initiates starting from epithelial cells and innate lymphoid cells (ILCs), including dendritic cells and macrophages, and translates to adaptive T and B-cell immunity, particularly IgE antibody production. Eosinophils, mast cells and basophils have effects on effector functions. Cytokines from ILC2s and CD4+ helper type 2 (Th2) cells, CD8 + T cells, and NK-T cells, along with myeloid cells, including IL-4, IL-5, IL-9, and IL-13, initiate and sustain allergic inflammation via T cell cells, eosinophils, and ILC2s; promote IgE class switching; and open the epithelial barrier. Epithelial cell activation, alarmin release and barrier dysfunction are key in the development of not only allergic diseases but also many other systemic diseases. Recent biologics targeting the pathways and effector functions of IL4/IL13, IL-5, and IgE have shown promising results for almost all ages, although some patients with severe allergic diseases do not respond to these therapies, highlighting the unmet need for a more detailed and personalized approach.

Iron homeostasis and ferroptosis in muscle diseases and disorders: mechanisms and therapeutic prospects

The muscular system plays a critical role in the human body by governing skeletal movement, cardiovascular function, and the activities of digestive organs. Additionally, muscle tissues serve an endocrine function by secreting myogenic cytokines, thereby regulating metabolism throughout the entire body. Maintaining muscle function requires iron homeostasis. Recent studies suggest that disruptions in iron metabolism and ferroptosis, a form of iron-dependent cell death, are essential contributors to the progression of a wide range of muscle diseases and disorders, including sarcopenia, cardiomyopathy, and amyotrophic lateral sclerosis. Thus, a comprehensive overview of the mechanisms regulating iron metabolism and ferroptosis in these conditions is crucial for identifying potential therapeutic targets and developing new strategies for disease treatment and/or prevention. This review aims to summarize recent advances in understanding the molecular mechanisms underlying ferroptosis in the context of muscle injury, as well as associated muscle diseases and disorders. Moreover, we discuss potential targets within the ferroptosis pathway and possible strategies for managing muscle disorders. Finally, we shed new light on current limitations and future prospects for therapeutic interventions targeting ferroptosis.

The antitumor activity of TGFβ-specific T cells is dependent on IL-6 signaling

Although interleukin (IL)-6 is considered immunosuppressive and tumor-promoting, emerging evidence suggests that it may support antitumor immunity. While combining immune checkpoint inhibitors (ICIs) and radiotherapy in patients with pancreatic cancer (PC) has yielded promising clinical results, the addition of an anti-IL-6 receptor (IL-6R) antibody has failed to elicit clinical benefits. Notably, a robust TGFβ-specific immune response at baseline in PC patients treated solely with ICIs and radiotherapy correlated with improved survival. Recent preclinical studies demonstrated the efficacy of a TGFβ-based immune modulatory vaccine in controlling PC tumor growth, underscoring the important role of TGFβ-specific immunity in PC. Here, we explored the importance of IL-6 for TGFβ-specific immunity in PC. In a murine model of PC, coadministration of the TGFβ-based immune modulatory vaccine with an anti-IL-6R antibody rendered the vaccine ineffective. IL-6R blockade hampered the development of vaccine-induced T-cells and tumoral T-cell infiltration. Furthermore, it impaired the myeloid population, resulting in increased tumor-associated macrophage infiltration and an enhanced immunosuppressive phenotype. In PC patients, in contrast to those receiving only ICIs and radiotherapy, robust TGFβ-specific T-cell responses at baseline did not correlate with improved survival in patients receiving ICIs, radiotherapy and IL-6R blockade. Peripheral blood immunophenotyping revealed that IL-6R blockade altered the T-cell and monocytic compartments, which was consistent with the findings in the murine model. Our data suggest that the antitumor efficacy of TGFβ-specific T cells in PC depends on the presence of IL-6 within the tumor. Consequently, caution should be exercised when employing IL-6R blockade in patients receiving cancer immunotherapy.

Engineering bone/cartilage organoids: strategy, progress, and application

The concept and development of bone/cartilage organoids are rapidly gaining momentum, providing opportunities for both fundamental and translational research in bone biology. Bone/cartilage organoids, essentially miniature bone/cartilage tissues grown in vitro, enable the study of complex cellular interactions, biological processes, and disease pathology in a representative and controlled environment. This review provides a comprehensive and up-to-date overview of the field, focusing on the strategies for bone/cartilage organoid construction strategies, progresses in the research, and potential applications. We delve into the significance of selecting appropriate cells, matrix gels, cytokines/inducers, and construction techniques. Moreover, we explore the role of bone/cartilage organoids in advancing our understanding of bone/cartilage reconstruction, disease modeling, drug screening, disease prevention, and treatment strategies. While acknowledging the potential of these organoids, we discuss the inherent challenges and limitations in the field and propose potential solutions, including the use of bioprinting for organoid induction, AI for improved screening processes, and the exploration of assembloids for more complex, multicellular bone/cartilage organoids models. We believe that with continuous refinement and standardization, bone/cartilage organoids can profoundly impact patient-specific therapeutic interventions and lead the way in regenerative medicine.

Targeting of TAMs: can we be more clever than cancer cells?

With increasing incidence and geography, cancer is one of the leading causes of death, reduced quality of life and disability worldwide. Principal progress in the development of new anticancer therapies, in improving the efficiency of immunotherapeutic tools, and in the personification of conventional therapies needs to consider cancer-specific and patient-specific programming of innate immunity. Intratumoral TAMs and their precursors, resident macrophages and monocytes, are principal regulators of tumor progression and therapy resistance. Our review summarizes the accumulated evidence for the subpopulations of TAMs and their increasing number of biomarkers, indicating their predictive value for the clinical parameters of carcinogenesis and therapy resistance, with a focus on solid cancers of non-infectious etiology. We present the state-of-the-art knowledge about the tumor-supporting functions of TAMs at all stages of tumor progression and highlight biomarkers, recently identified by single-cell and spatial analytical methods, that discriminate between tumor-promoting and tumor-inhibiting TAMs, where both subtypes express a combination of prototype M1 and M2 genes. Our review focuses on novel mechanisms involved in the crosstalk among epigenetic, signaling, transcriptional and metabolic pathways in TAMs. Particular attention has been given to the recently identified link between cancer cell metabolism and the epigenetic programming of TAMs by histone lactylation, which can be responsible for the unlimited protumoral programming of TAMs. Finally, we explain how TAMs interfere with currently used anticancer therapeutics and summarize the most advanced data from clinical trials, which we divide into four categories: inhibition of TAM survival and differentiation, inhibition of monocyte/TAM recruitment into tumors, functional reprogramming of TAMs, and genetic enhancement of macrophages.

Responses

Your email address will not be published. Required fields are marked *